Chinese Journal of Endocrinology and Metabolism ; (12): 634-635, 2011.
Article in Chinese | WPRIM | ID: wpr-424265

ABSTRACT

Blood samples were detected by 24 blood-glucose meters using glucose oxidase and another 18 glucose meters using glucose dehydrogenase. The plasma glucose was detected by the auto-chemistry analyzer as control. According to ISO1 5197 and EF9-A2, the bias of results from both glucose meters 100% fall in the range of ±0. 56 mmol/L( <4. 2 mmol/L) and ≥99. 7% in the range of ±20% ( ≥4. 2 mmol/L), predicted bias were all less than the true bias(Bc). As κ>0. 6, the results from both glucose meters were in accordance with the results from autochemistry analyzer in judging hyperglycemia ( >6. 11 mmol/L) and hypoglycemia ( < 3.89 mmol/L). Between the results from the two blood-glucose meters, κ<0. 6. The accuracy of both glucose-meters are accepted for the purpose of clinical diagnosis and treatment.